Sensor: 123mm(L)*22mm(W)*19.5mm(H)
Main Unit: 110mm(L)*80mm(W)*27mm(H)
The KD5002 Electrostatic Field Monitor (hereinafter referred to as "the Monitor") is a non-contact electrostatic field monitoring instrument developed by our company for surface process monitoring in the electronics manufacturing industry. Its sensitive probe utilizes an electrostatic sensitive element based on MEMS technology to measure surface electrostatic field strength through a non-contact method.
The shielding electrode of the sensor is grounded. Driven by an excitation voltage, the excitation electrode causes the shielding electrode to vibrate horizontally at a frequency of ω, periodically shielding the sensing electrode. This results in a periodic change in the induced charge on the surface of the sensing electrode, thereby generating an induced current. The amplitude of this current is directly proportional to the amplitude of the measured electric field, allowing the target electric field to be determined by measuring this current value. Because the instrument contains no motors or other easily worn mechanical components, it offers superior stability and higher reliability.
Equipped with a dedicated communication converter, the Monitor acquires real-time detection data and displays the instantaneous electrostatic field values. It can transmit and store data to a computer via the ZigBee protocol. Using the accompanying software, users can view real-time measurement data and perform comprehensive data analysis on their computers.
| Product | ESD Field Monitor |
Model | KD5002 |
Monitoring Target | Electrostatic Field |
| Monitoring Range (V) | -10KV~+10KV |
| Outer Cover | SGCC painted |
| Usage Recommendations | 1. Keep the sensor probe as close as possible to the object under test; 5 cm is optimal. Other distances can also be calibrated online. |
| 2. Keep the probe away from equipment with built-in high-voltage discharge units (e.g., CRT monitors, SMT soldering machines, electrostatic eliminators, etc.). |
* The above information is for reference only, subject to corresponding specifications.